Comparison of Postoperative Coronal Leg Alignment in Customized Individually Made and Conventional Total Knee Arthroplasty

2021

Neutral coronal leg alignment is known to be important for postoperative outcome in total knee arthroplasty (TKA). Customized individually made implants (CIM) instrumented with patient-specific cutting guides are an innovation aiming to increase the precision and reliability of implant positioning and reconstruction of leg alignment. We aimed to compare reconstruction of the hip–knee–ankle angle (HKA) of the novel CIM system iTotal™ CR G2 (ConforMIS Inc.) to a matched cohort of the off-the-shelf (OTS) knee replacement system Vanguard™ CR (Zimmer Biomet). Retrospective analysis of postoperative coronal full-leg weight-bearing radiographs of 562 TKA (283 CIM TKA, 279 OTS TKA) was conducted. V…

musculoskeletal diseasestotal knee arthroplastyleg alignmentRadiographymedicine.medical_treatmentTotal knee arthroplastyMedicine (miscellaneous)Knee replacementArticlepatient-specific instruments03 medical and health sciences0302 clinical medicineMedicinePostoperative outcomeOrthodontics030222 orthopedicsRotational errorbiologybusiness.industryR030229 sport sciencesmusculoskeletal systembiology.organism_classificationValguscustom-made implantCoronal planeMedicinerotational correctionImplantbusinessJournal of Personalized Medicine
researchProduct